reverse osmosis system

    reverse osmosis system

    2200 Reverse Osmosis System

    Up to 2,200 GPD

    Pure water used for acute hemodialysis must be of the same quality as water used in chronic settings. However, a larger water system is often not affordable or practical. Single-patient systems can prove costly if several are necessary. Therefore, central acute services require a complete, compact, and economical system that can produce high-quality water to meet the demand for dialyzing as many as six patient.

     

     

     

Standard Features
• Digital water-quality meter (LCD screen displays product and incoming water quality, percent rejection, and activates audible and visual alarms.
• Standard permeate divert function on the RO.
• Adjustable permeate relief valve to drain.
• Panel-mount pressure gauges and flow meters on every machine
• Stainless steel RO pump for long-lasting, quiet operation and including a soft start feature.
• Swivel locking casters that provide easy movement.
• RO can be setup for Direct or Indirect feed.
• Pretreatment lock-out and remote alarm interface options.
• Availability of installation kits for quick and easy installation.

 

 

 

 

2200 Specifications
Part Number100-32-562 
Membrane(2) 4 x 21 TFC 
Production @ 77o F (25o C)5800 cc/ml min (1.52 gpm)
2200 Specifications  
Electrical115 VAC, 20 AMP, 60 Hz circuit, Single Phase plug-in
Connections3/4” MGHT fittings
Feed Water Requirements  Temperature: 40-90o F (4-32 o C)Total Chlorine: < 0.0 ppm
Inlet Pressure: 20-60 psigPrefiltration: <1 μm
pH: 5-10SDI: < 3
Dimensions12.5” W x 39” H x 28” D
MaterialsBrass, Stainless steel or inert plastic
Performance> 99% ionic rejection with properly pretreated feed water
Recovery~50% to 75% 
Weight140 lbs. (63.6 Kg)